Page images
PDF
EPUB

graph (c) of this section, analyses for total trihalomethanes shall be performed at quarterly intervals on at least four water samples for each treatment plant used by the system. At least 25 percent of the samples shall be taken at locations within the distribution system reflecting the maximum residence time of the water in the system. The remaining 75 percent shall be taken at representative locations in the distribution system, taking into account number of persons served, different sources of water and different treatment methods employed. The results of all analyses per quarter shall be arithmetically averaged and reported to the State within 30 days of the system's receipt of such results. Results shall also be reported to EPA until such monitoring requirements have been adopted by the State. All samples collected shall be used in the computation of the average, unless the analytical results are invalidated for technical reasons. Sampling and analyses shall be conducted in accordance with the methods listed in paragraph (e) of this section.

(2) Upon the written request of a community water system, the monitoring frequency required by paragraph (b)(1) of this section may be reduced by the State to a minimum of one sample analyzed for TTHMs per quarter taken at a point in the distribution system reflecting the maximum residence time of the water in the system, upon a written determination by the State that the data from at least 1 year of monitoring in accordance with paragraph (b)(1) of this section and local conditions demonstrate that total trihalomethane concentrations will be consistently below the maximum contaminant level.

(3) If at any time during which the reduced monitoring frequency prescribed under this paragraph applies, the results from any analysis exceed 0.10 mg/l of TTHMS and such results are confirmed by at least one check sample taken promptly after such results are received, or if the system makes any significant change to its source of water or treatment program, the system shall immediately begin monitoring in accordance with the requirements of paragraph (b)(1) of this

section, which monitoring shall continue for at least 1 year before the frequency may be reduced again. At the option of the State, a system's monitoring frequency may and should be increased above the minimum in those cases where it is necessary to detect variations of TTHM levels within the distribution system.

(c)(1) Upon written request to the State, a community water system utilizing only ground water sources may seek to have the monitoring frequency required by paragraph (b)(1) of this section reduced to a minimum of one sample for maximum TTHM potential per year for each treatment plant used by the system taken at a point in the distribution system reflecting maximum residence time of the water in the system. The system shall submit to the State the results of at least one sample analyzed for maximum TTHM potential for each treatment plant used by the system taken at a point in the distribution system reflecting the maximum residence time of the water in the system. The system's monitoring frequency may only be reduced upon a written determination by the State that, based upon the data submitted by the system, the system has a maximum TTHM potential of less than 0.10 mg/l and that, based upon an assessment of the local conditions of the system, the system is not likely to approach or exceed the maximum contaminant level for total TTHMS. The results of all analyses shall be reported to the State within 30 days of the system's receipt of such results. Results shall also be reported to EPA until such monitoring requirements have been adopted by the State. All samples collected shall be used for determining whether the system must comply with the monitoring requirements of paragraph (b) of this section, unless the analytical results are invalidated for technical reasons. Sampling and analyses shall be conducted in accordance with the methods listed in paragraph (e) of this section.

(2) If at any time during which the reduced monitoring frequency prescribed under paragraph (c)(1) of this section applies, the results from any analysis taken by the system for maximum TTHM potential are equal to or

greater than 0.10 mg/l, and such results are confirmed by at least one check sample taken promptly after such results are received, the system shall immediately begin monitoring in accordance with the requirements of paragraph (b) of this section and such monitoring shall continue for at least one year before the frequency may be reduced again. In the event of any significant change to the system's raw water or treatment program, the system shall immediately analyze an additional sample for maximum TTHM potential taken at a point in the distribution system reflecting maximum residence time of the water in the system for the purpose of determining whether the system must comply with the monitoring requirements of paragraph (b) of this section. At the option of the State, monitoring frequencies may and should be increased above the minimum in those cases where this is necessary to detect variation of TTHM levels within the distribution system.

(d) Compliance with § 141.12(c) shall be determined based on a running annual average of quarterly samples collected by the system as prescribed in paragraph (b)(1) or (2) of this section. If the average of samples covering any 12 month period exceeds the Maximum Contaminant Level, the supplier of water shall report to the State pursuant to § 141.31 and notify the public pursuant to § 141.32. Monitoring after public notification shall be at a frequency designated by the State and shall continue until a monitoring schedule as a condition to a variance, exemption or enforcement action shall become effective.

(e) Sampling and analyses made pursuant to this section shall be conducted by one of the following EPA approved methods:

(1) "The Analysis of Trihalomethanes in Drinking Waters by the Purge and Trap Method," Method 501.1, EMSL, EPA Cincinnati, Ohio.

(2) "The Analysis of Trihalomethanes in Drinking Water by Liquid/ Liquid Extraction," Method 501.2, EMSL, EPA Cincinnati, Ohio. Samples for TTHM shall be dechlorinated upon collection to prevent further production of Trihalomethanes,

according to the procedures described in the above two methods. Samples for maximum TTHM potential should not be dechlorinated, and should be held for seven days at 25° C (or above) prior to analysis, according to the procedures described in the above two methods.

(f) Before

a community water

system makes any significant modifications to its existing treatment process for the purposes of achieving compliance with § 141.12(c), such system must submit and obtain State approval of a detailed plan setting forth its proposed modification and those safeguards that it will implement to ensure that the bacteriological quality of the drinking water served by such system will not be adversely affected by such modification. Each system shall comply with the provisions set forth in the State-approved plan. At a minimum, a State approved plan shall require the system modifying its disinfection practice to:

(1) Evaluate the water system for sanitary defects and evaluate the source water for biological quality;

(2) Evaluate its existing treatment practices and consider improvements that will minimize disinfectant demand and optimize finished water quality throughout the distribution system;

(3) Provide baseline water quality survey data of the distribution system. Such data should include the results from monitoring for coliform and fecal coliform bacteria, fecal streptococci, standard plate counts at 35° C and 20° C, phosphate, ammonia nitrogen and total organic carbon. Virus studies should be required where source waters are heavily contaminated with sewage effluent;

(4) Conduct additional monitoring to assure continued maintenance of optimal biological quality in finished water, for example, when chloramines are introduced as disinfectants or when pre-chlorination is being discontinued. Additional monitoring should also be required by the State for chlorate, chlorite and chlorine dioxide when chlorine dioxide is used. Standard plate count analyses should also be required by the State as appropriate before and after any modifications;

(5) Consider inclusion in the plan of provisions to maintain an active disinfectant residual throughout the distribution system at all times during and after the modification.

[44 FR 68641, Nov. 29, 1979, as amended at 45 FR 15545, 15547, Mar. 11, 1980]

APPENDIX A—SUMMARY OF PUBLIC COMMENTS AND EPA RESPONSES ON PROPOSED AMENDMENTS TO THE NATIONAL INTERIM PRIMARY DRINKING WATER REGULATIONS FOR CONTROL OF TRIHALOMETHANES IN DRINKING WATER

[44 FR 68642, Nov. 29, 1979]

APPENDIX B-SUMMARY OF MAJOR COMMENTS (FOR RESPONSES, SEE APPENDIX A)

[44 FR 68666, Nov. 29, 1979]

EDITORIAL NOTE: Appendices A and B appearing at 44 FR 68642 and 68666, Nov. 29, 1979, are not codified in the Code of Federal Regulations.

APPENDIX C-ANALYSIS OF
TRIHALOMETHANES

PART I-THE ANALYSIS OF TRIHALOMETHANES
IN DRINKING WATER BY THE PURGE AND
TRAP METHOD

1. Scope

1.1 This method (1) is applicable in the determination of four trihalomethanes, i.e. chloroform, dichlorobromomethane, dibromochloromethane, and bromoform in finished drinking water, raw source water, or drinking water in any stage of treatment. The concentration of these four compounds is totaled to determine total trihalomethanes (TTHM).

1.2 For compounds other than the abovementioned trihalomethanes, or for other sample sources, the analyst must demonstrate the usefulness of the method by collecting precision and accuracy data on actual samples as described (2).

1.3 Although the actual detection limits are highly dependent upon the gas chromatographic column and detector employed, the method can be used over a concentration range of approximately 0.5 to 1500 micrograms per liter.

1.4 Well in excess of 100 different water supplies have been analyzed using this method. Supplementary analyses using gas chromatography mass spectrometry (GC/ MS) have shown that there is no evidence of interference in the determination of trihalo

methanes (3). For this reason, it is not necessary to analyze the raw source water as is required with the Liquid/Liquid Extraction Method (4).

2. Summary

2.2 Trihalomethanes are extracted by an inert gas which is bubbled through the aqueous sample. The trihalomethanes, along with other organic constituents which exhibit low water solubility and a vapor pressure significantly greater than water, are efficiently transferred from the aqueous phase to the gaseous phase. These compounds are swept from the purging device and are trapped in a short column containing a suitable sorbent. After a predetermined period of time, the trapped components are thermally desorbed and backflushed onto the head of a gas chromatographic column and separated under programmed conditions. Measurement is accomplished with a halogen specific detector such as electrolytic conductivity or microcoulometric titration.

2.3 Confirmatory analyses are performed using dissimilar columns, or by mass spectrometry (5).

2.4 Aqueous standards and unknowns are extracted and analyzed under identical conditions in order to compensate for extraction losses.

2.5 The total analysis time, assuming the absence of other organohalides, is approximately 35 minutes per sample.

3. Interferences

3.1 Impurities contained in the purge gas and organic compounds outgasing from the plumbing ahead of the trap usually account for the majority of contamination problems. The presence of such inteferences are easily monitored as a part of the quality control program. Sample blanks are normally run between each set of samples. When a positive trihalomethane response is noted in the sample blank, the analyst should analyze a method blank. Method blanks are run by charging the purging device with organicfree water and analyzing in the normal

manner.

If any trihalomethane is noted in the method blank in excess of 0.4 μg/l, the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter. Subtracting the blank values is not recommended. The use of non-TFE plastic tubing, non-TFE thread sealants, or flow controllers with rubber components should be avoided since such materials generally out-gas organic compounds which will be concentrated in the trap during the purge operation. Such out-gasing problems are common whenever new equipment is put

into service; as time progresses, minor outgasing problems generally cure themselves.

3.2 Several instances of accidental sample contamination have been noted and attributed to diffusion of volatile organics through the septum seal and into the sample during shipment and storage. The sample blank is used as a monitor for this problem.

3.3 For compounds that are not efficiently purged, such as bromoform, small variations in sample volume, purge time, purge flow rate, or purge temperature can affect the analytical result. Therefore, samples and standards must be analyzed under identical conditions.

3.4 Cross-contamination can occur whenever high-level and low-level samples are sequentially analyzed. To reduce this likelihood, the purging device and sample syringe should be rinsed twice between samples with organic-free water. Whenever an unusually concentrated sample is encountered, it is highly recommended that it be followed by a sample blank analysis to ensure that sample cross contamination does not occur. For samples containing large amounts of water soluble materials, it may be necessary to wash out the purging device with a soap solution, rinse with distilled water, and then dry in a 105°C oven between analyses.

3.5 Qualitative misidentifications are a problem in using gas chromatographic analysis. Whenever samples whose qualitative nature is unknown are analyzed, the following precautionary measures should be incorporated into the analysis.

3.5.1 Perform duplicate analyses using the two recommended columns (4.2.1 and 4.2.2) which provide different retention order and retention times for the trihalomethanes and other organohalides.

3.5.2 Whenever possible, use GC/MS techniques which provide unequivocal qualitative identifications (5).

[blocks in formation]

space for most foams to disperse. The inlet and exit ports are constructed from heavywalled 1⁄4-inch glass tubing so that leak-free removable connections can be made using "finger-tight” compression fittings containing Teflon ferrules. The removable foam trap is used to control samples that foam.

4.1.2 Trapping Device-The trap (Figure 2) is a short gas chromatographic column which at <35° C retards the flow of the compounds of interest while venting the purge gas and, depending on which sorbent is used, much of the water vapor. The trap should be constructed with a low thermal mass so that it can be heated to 180° C in less than 1 minute for efficient desorption, then rapidly cooled to room temperature for recycling. Variations in the trap ID, wall thickness, sorbents, sorbent packing order, and sorbent mass could adversely affect the trapping and desorption efficiencies for compounds discussed in this text. For this reason, it is important to faithfully reproduce the trap configurations recommended in Figure 2. Traps containing Tenax only, or combinations of Tenax and other sorbents are acceptable for this analysis.

4.1.3 Desorber assembly-Details for the desorber are shown in Figures 3, and 4. With the 6-port valve in the Purge Sorb position (Figure 3), the effluent from the purging device passes through the trap where the flow rate of the organics is retarded. The GC carrier gas also passes through the 6-port valve and is returned to the GC. With the 6-port valve in the PurgeSorb position, the operation of the GC is in no way impaired; therefore, routine liquid injection analyses can be performed using the gas chromatograph. After the sample has been purged, the 6-port valve is turned to the desorb position (Figure 4). In this configuration the trap is coupled in series with the gas chromatographic column allowing the carrier gas to backflush the trapped materials into the analytical column. Just as the valve is actuated, the power is turned on to the resistance wire wrapped around the trap. The power is supplied by an electronic temperature controller. Using this device, the trap is rapidly heated to 180° C and then maintained at 180° C with minimal temperature overshoot. The trapped compounds are released as a "plug" to the gas chromatograph. Normally, packed columns with theoretical efficiencies near 500 plates/foot under programmed temperature conditions can accept such desorb injections without altering peak geometry. Substituting a non-controlled power supply, such as a manually-operated variable transformer, will provide nonreproductible retention times and poor quantitative data unless Injection Procedure (8.9.2) is used.

4.1.4 Several Purge and Trap Devices are now commercially available. It is recommended that the following be taken into consideration if a unit is to be purchased:

a. Be sure that the unit is completely compatible with the gas chromatograph to be used for the analysis.

b. Use a 5-ml purging device similar to that shown in Figure 1.

c. Be sure the Tenax portion of the trap meets or exceeds the dimensions shown in Figure 2.

d. With the exception of sample introduction, select a unit that has as many of the purge trap functions automated as possible. 4.2 Gas chromatograph-The chromatograph must be temperature programmable and equipped with a halide specific detector.

4.2.1 Column I is an unusually efficient column which provides outstanding separations for a wide variety of organic compounds. Because of its ability to resolve trihalomethanes from other organochlorine compounds, column I should be used as the primary analytical column (see Table 1 for retention data using this column).

4.2.1.1 Column I parameters: Dimensions-8 feet long x 0.1 inch ID stainless steel or glass tubing. Packing-1% SP-1000 on Carbopack-B (60/80) mesh. Carrier Gas-helium at 40 ml/minute. Temperature program sequence: 45° C isothermal for 3 minutes, program at 8° C/minute to 220° C then hold for 15 minutes or until all compounds have eluted.

NOTE: It has been found that during handling, packing, and programming, active sites are exposed on the Carbopack-B packing. This results in tailing peak geometry and poor resolution of many constituents. To correct this, pack the first 5 cm of the column with 3% SP-1000 on Chromosorb-W 60/80 followed by the Carbopack-B packing. Condition the precolumn and the Carbopack columns with carrier gas flow at 220° C overnight. Pneumatic shocks and rough treatment of packed columns will cause excessive fracturing of the Carbopack. If pressure in excess of 60 psi is required to obtain 40 ml/minute carrier flow, then the column should be repacked.

4.2.1.2 Acceptable column equivalent to Column I: Dimensions-8 feet long x 0.1 inch ID stainless steel or glass tubing. Packing-0.2% Carbowax 1500 on Carbopack-C (80/100) mesh. Carrier Gas-helium at 40 ml/minute. Temperature program sequence-60° C isothermal for 3 minutes, program at 8° C /minute to 160° C, then hold for 2 minutes or until all compounds have eluted.

NOTE: It has been found that during handling, packing, and programming, active sites are exposed on the Carbopack-C packing. This results in poor resolution of con

stituents and poor peak geometry. To correct this, place a 1 ft. 0.125 in. OD x 0.1 in. ID stainless steel column packed with 3% Carbowax 1500 on Chromosorb-W 60/80 mesh in series before the Carbopack-C column. Condition the precolumn and the Carbopack columns with carrier gas flow at 190° C overnight. The two columns may be retained in series for routine analyses. Trihalomethane retention times are listed in Table 1.

4.2.2 Column II provides unique organohalide-trihalomethane separations when compared to those obtained from Column I (see Figures 5 and 6). However, since the resolution between various compounds is generally not as good as those with Column I, it is recommended that Column II be used as a qualitative confirmatory column for unknown samples when GC/MS confirmation is not possible.

4.2.2.1 Column II parameters: Dimensions-6 feet long x 0.1 inch ID stainless steel or glass. Packing-n-octane on PorisilC (100/120 mesh). Carrier Gas-helium at 40 cc/minute. Temperature program sequence-50° C isothermal for 3 minutes, program at 6°/minute to 170° C, then hold for 4 minutes or until all compounds have eluted. Trihalomethane retention times are listed in Table 1.

5.8 Organic-free water is defined as water free of interference when employed in the purge and trap analysis.

5.8.1 Organic-free water is generated by passing tap water through a carbon filter bed containing about 1 lb. of activated carbon. Change the activated carbon bed whenever the concentration of any trihalomethane exceeds 0.4 μg/l.

5.8.2 A Millipore Super-Q Water System or its equivalent may be used to generate organic-free water.

5.8.3 Organic-free water may also be prepared by boiling water for 15 minutes. Subsequently, while maintaining the temperature at 90° C, bubble a contaminant-free inert gas through the water for one hour. While still hot, transfer the water to a narrow-mouth screw-cap bottle with a Teflon seal.

5.8.4 Test organic free water each day it is used by analyzing according to Section 8. 5.9 Standards.⚫

[blocks in formation]
« PreviousContinue »