E Lift bracket al center of generator and remove generator from mounting hal. E Generator is spring-loaded and could pop out if not held in position when center bracket is lifted. F Remove oxygen supply hoses from generator oxygen outlets. F. G. Coil firing pin lanyards and secure to generator body. 16 Install Generator. A. Clear lanyards from generator body. CAUTION: DO NOT USE ANY LUBRICANT TO EASE HOSE ONTO OUTLET. IF ANY EVIDENCE INDICATES 2- A to D. LUBRICANT WAS PREVIOUSLY USED, WIPE OFF OUTLET AND TRIM ABOUT ONE INCH (25.4 Connect oxygen supply hoses to generator oxygen oullels and check that hoses are installed at least Lift center bracket and engage generator side having two pins into holes in mounting hat, then lower center CAUTION: MINIMUM 5/8-INCH (15.9 MM) CLEARANCE MUST BE MAINTAINED TO ALLOW PROPER Insert heat shield bracket lugs into mounting holes in bottom of oxygen module. Check that 5/8-inch (15.9 Note: Heat shield is not incorporated in forward attendant's module. CAUTION: USE CAUTION NOT TO PULL GENERATOR FIRING PIN WHILE CONNECTING LANYARDS. E. Connect firing pin lanyards to masks using bowline knot installation (Figures 1 and 2). With module door in open position, make certain that lanyard length is as follows: 69.00(:10.50) inches (1752.6(:1:12.7) mm) mea: ring H- 12.00(10.50) inches (304.8(12.7) mm) measuring 50(:1:1.00) inches (1270.0(±25.4) mm) measuring from 1.00(:10.50) Inch (25.4(112.7 mm) slack F Remove excess lanyard. Note: Reset door latching mechanism by placing release lever flush with latching coil. As door is closed, the door mounted latch strike will engage strike retainer and will be held in pie by an internal spring-loaded mechanism. Remove safely cap from generator primer and check that firing mechanism of th cooked position. oxygen generator is in A/C NO. N803VY MECII SIGNOFF 712 Close oxygen module door slowly. ALL ITEMS SIGNED EMP DATE 05/1564-28-9 McDonnell Douglas Corporation Proprietary information Use or disclosure of |