Page images
PDF
EPUB
[blocks in formation]
[blocks in formation]

PART II-ANALYSIS OF TRIHALOMETHANES IN DRINKING WATER BY LIQUID/LIQUID EXTRACTION

1. Scope.

1.1 This method (1,2) is applicable only to the determination of four trihalomethanes, i.e., chloroform, bromodichloromethane, chlorodibromomethane, and bromoform in finished drinking water, drinking water during intermediate stages of treatment, and the raw source water.

1.2 For compounds other than the abovementioned trihalomethanes, or for other sample sources, the analyst must demonstrate the usefulness of the method by collecting precision and accuracy data on actual samples as described in (3) and provide qualitative confirmation of results by Gas Chromatography/Mass Spectrometry (GC/MS) (4).

1.3 Qualitative analyses using GC/MS or the purge and trap method (5) must be performed to characterize each raw source water if peaks appear as interferences in the raw source analysis.

1.4 The method has been shown to be useful for the trihalomethanes over a concentration range from approximately 0.5 to 200 μg/l. Actual detection limits are highly dependent upon the characteristics of the gas chromatographic system used.

2. Summary

2.1 Ten milliliters of sample are extracted one time with 2 ml of solvent. Three μl of the extract are then injected into a gas chromatograph equipped with a linearized electron capture detector for separation and analysis.

2.2 The extraction and analysis time is 10 to 50 minutes per sample depending upon the analytical conditions chosen. (See Table 1 and Figures 1, 2, and 3.)

2.3 Confirmatory evidence is obtained using dissimilar columns and temperature programming. When component concentrations are sufficiently high (>50 μg/l), halogen specific detectors may be employed for improved specificity.

2.4 Unequivocal confirmatory analyses at high levels (>50 μg/l) can be performed using GC/MS in place of the electron capture detector. At levels below 50 μg/l, unequivocal confirmation can only be performed by the purge and trap technique using GC/MS (4, 5).

2.5 Standards dosed into organic free water and the samples are extracted and analyzed in an identical manner in order to compensate for possible extraction losses.

2.6 The concentration of each trihalomethane is summed and reported as total trihalomethanes in μg/l.

3. Interferences

3.1 Impurities contained in the extracting solvent usually account for the majority of the analytical problems. Solvent blanks should be analyzed before a new bottle of solvent is used to extract samples. Indirect daily checks on the extracting solvent are obtained by monitoring the sample blanks (6.4.10). Whenever an interference is noted in the sample blank, the analyst should reanalyze the extracting solvent. The extraction solvent should be discarded whenever a high level (>10 μg/l) of interfering compounds are traced to it. Low level interferences generally can be removed by distillation or column chromatography (6); however, it is generally more economical to obtain a new source of solvent or select one of the approved alternative solvents listed in Section 5.1. Interference free solvent is defined as a solvent containing less than 0.4 μg/l individual trihalomethane interference. Protect interference-free solvents by storing in a non-laboratory area known to be free of organochlorine solvents. Subtracting blank values is not recommended.

3.2 Several instances of accidental sample contamination have been attributed to diffusion of volatile organics through the septum seal on the sample bottle during shipment and storage. The sample blank (6.4.10) is used to monitor for this problem.

3.3 This liquid/liquid extraction technique efficiently extracts a wide boiling range of non-polar organic compounds and, in addition, extracts the polar organic components of the sample with varying efficiencies. In order to perform the trihalomethane analysis as rapidly as possible with sensitivities in the low μg/l range, it is necessary to use the semi-specific electron capture detector and chromatographic columns which have relatively poor resolving power. Because of these concessions, the probability of experiencing chromatographic interferences is high. Trihalomethanes are primarily products of the chlorination process and generally do not appear in the raw source water. The absence of peaks in the raw source water analysis with retention times similar to the trihalomethanes is generally adequate evidence of an interferencefree finished drinking water analysis. Because of these possible interferences, in addition to each finished drinking water analysis, a representative raw source water (6.4.5) must be analyzed. When potential interferences are noted in the raw source water analysis, the alternate chromatographic columns must be used to reanalyze the sample set. If interferences are still noted, qualitative identifications should be performed according to Sections 2.3 and 2.4. If the peaks are confirmed to be other than trihalomethanes and add significantly to

the total trihalomethane value in the finished drinking water analysis, then the sample set must be analyzed by the purge and trap method (5).

4. Apparatus

4.1 Extraction vessel-A 15 ml total volume glass vessel with a Teflon lined screw-cap is required to efficiently extract the samples.

4.1.1 For samples that do not form emulsions 10 ml screw-cap flasks with a Teflon faced septum (total volume is ml) are recommended. Flasks and caps-Pierce#13310 or equivalent. Septa-Teflon silicone-Pierce #12718 or equivalent.

4.1.2 For samples that form emulsions (turbid source water) 15 ml screw cap centrifuge tubes with a Teflon cap liner are recommended. Centrifuge tube-Corning 806215 or equivalent.

4.2 Sampling containers-40 ml screw cap sealed with Teflon faced silicone septa. Vials and caps-Pierce #13075 or equivalent. Septa-Pierce #12722 or equivalent.

4.3 Micro syringes-10, 100 μl.

4.4 Micro syringe-25 μl with a 2-inch by 0.006-inch needle-Hamilton 702N or equivalent.

4.5 Syringes-10 ml glass hypodermic with luerlok tip (2 each).

4.6 Syringe valve-2-way with luer ends (2 each)-Hamilton #86570-1FM1 or equivalent.

4.7 Pipette-2.0 ml transfer.

4.8 Glass stoppered volumetric flasks-10 and 100 ml.

4.9 Gas chromatograph with linearized electron capture detector. (Recommended option-temperature programmable. See Section 4.12.)

4.10 Column A-4 mm ID x 2m long glass packed with 3% SP-1000 on Supelcoport (100/120 mesh) operated at 50°C with 60 ml/min flow. (See Figure 1 for a sample chromatogram and Table 1 for retention

data.)

4.11 Column B-2 mm ID x 2m long glass packed with 10% squalane on Chromosorb WAW (80/100 mesh) operated at 67°C with 25 ml/min flow. This column is recommended as the primary analytical column. Trichloroethylene, a common raw source water contaminate, coelutes with bromodichloromethane. (See Figure 2 for a sample chromatogram and Table 1 for retention data.)

4.12 Column C-2 mm ID x 3m long glass packed with 6% OV-11/4% SP-2100 on Supelcoport (100/120 mesh) temperature program 45°C for 12 minutes, then program at 1°/minute to 70°C with a 25 ml/min flow. (See Figure 3 for a sample chromatogram and Table I for retention data.)

4.13 Standard storage containers-15 ml amber screw-cap septum bottles with Teflon faced silicone septa. Bottles and caps

[blocks in formation]

5.6 Organic-free

water-Organic-free water is defined as water free of interference when employed in the procedure described herein.

5.6.1 Organic-free water is generated by passing tap water through a carbon filter bed containing carbon. Change the activated carbon whenever the concentration of any trihalomethane exceeds 0.4 μg/l.

5.6.2 A Millipore Super-Q Water System or its equivalent may be used to generate organic-free deionized water.

5.6.3 Organic-free water may also be prepared by boiling water for 15 minutes. Subsequently, while maintaining the temperature at 90° C, bubble a contaminant free inert gas through the water at 100 ml/ minute for one hour. While still hot, transfer the water to a narrow mouth screw cap bottle with a Teflon seal.

Pentane has been selected as the best solvent for this analysis because it elutes, on all of the columns, well before any of the trihalomethanes. High altitudes or laboratory temperatures in excess of 75°F may make the use of this solvent impractical. For these reasons, alternative solvents are acceptable; however, the analyst may experience baseline variances in the elution areas of the trihalomethanes due to coelution of these solvents. The degree of difficulty appears to be dependent upon the design and condition of the electron capture detector. Such problems should be insignificant when concentrations of the coeluting trihalomethane are in excess of 5 μg/1.

As a precautionary measure, all standards must be checked for purity by boiling point determinations or GC/MS assays.

5.6.4 Test organic free water each day it is used by analyzing it according to Section 7.

5.7 Standard stock solutions.

5.7.1 Fill a 10.0 ml ground glass stoppered volumetric flask with approximately 9.8 ml of methyl alcohol.

5.7.2 Allow the flask to stand unstoppered about 10 minutes or until all alcohol wetted surfaces dry.

5.7.3 Weigh the unstoppered flask to the nearest 0.1 mg.

5.7.4 Using a 100 μl syringe, immediately add 2 to 3 drops of the reference standard to the flask, then reweigh. Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask.

5.7.5 Dilute to volume, stopper, then mix by inverting the flask several times.

5.7.6 Transfer the standard solution to a dated and labeled 15 ml screw-cap bottle with a Teflon cap liner.

NOTE: Because of the toxicity of trihalomethanes, it is necessary to prepare primary dilutions in a hood. It is further recommended that a NIOSH/MESA-approved toxic gas respirator be used when the analyst handles high concentrations of such materials.

5.7.7 Calculate the concentration in micrograms per microliter from the net gain in weight.

5.7.8 Store the solution at 4° C.

NOTE: All standard solutions prepared in methyl alcohol are stable up to 4 weeks when stored under these conditions. They should be discarded after that time has elapsed.

5.8 Aqueous calibration standard precautions.

5.8.1 In order to prepare accurate aqueous standard solutions, the following precautions must be observed:

a. Do not inject more than 20 μl of alcoholic standards into 100 ml of organic-free water.

b. Use a 25 μl Hamilton 702N microsyringe or equivalent. (Variations in needle geometry will adversely affect the ability to deliver reproducible volumes of methanolic standards into water.)

c. Rapidly inject the aloholic standard into the expanded area of the filled volumetric flask. Remove the needle as fast as possible after injection.

d. Mix aqueous standards by inverting the flask three times only.

e. Discard the contents contained in the neck of the flask. Fill the sample syringe from the standard solution contained in the expanded area of the flask as directed in Section 7.

f. Never use pipets to dilute or transfer samples and aqueous standards.

[blocks in formation]

5.9.1 Prepare, from the standard stock solutions, a multicomponent secondary dilution mixture in methyl alcohol so that a 20 μl injection into 100 ml of organic-free water will generate a calibration standard which produces a response close (± 25%) to that of the unknown. (See 8.1.)

5.9.2 Alternative calibration procedure. 5.9.2.1 Construct a calibration curve for each trihalomethane containing a minimum of 3 different concentrations. Two of the concentrations must bracket each unknown.

5.9.3 Extract and analyze the aqueous calibration standards in the same manner as the unknowns.

5.9.4 Other calibration procedures (7) which require the delivery of less than 20 μl of methanolic standards to 10.0 ml volumes of water contained in the sample syringe are acceptable only if the methanolic standard is delivered by the solvent flush technique (8).

5.10 Quality Check Standard Mixture.

5.10.1 Prepare, from the standard stock solutions, a secondary dilution mixture in methyl alcohol that contains 10.0 ng/μl of each compound. (See 5.7.6 and 5.7.8.)

5.10.2 Daily, prepare and analyze a 2.0 μg/l aqueous dilution from this mixture by dosing 20.0 μl into 100 ml of organic-free water (See Section 8.1).

6. Sample Collection and Handling. 6.1 The sample containers should have a total volume of at least 25 ml.

6.1.1 Narrow-mouth screw-cap bottles with the TFE fluorocarbon faced silicone septa cap liners are strongly recommended. 6.2 Glassware Preparation.

6.2.1 Wash all sample bottles, TFE seals, and extraction flasks in detergent. Rinse with tap water and finally with distilled water.

6.2.2 Allow the bottles and seals to air dry, then place in an 105° C oven for 1 hour, then allow to cool in an area known to be free of organics.

NOTE: Do not heat the TFE seals for extended periods of time (>1 hour) because the silicone layer slowly degrades at 105° C.

6.2.3 When cool, seal the bottles using the TFE seals that will be used for sealing the samples.

6.3 Sample stabilization-A chemical reducing agent (Section 5.3) is added to all samples in order to arrest the formation of additional trihalomethanes after sample collection (7,9) and to eliminate the possibility of free chlorine reacting with impurities

« PreviousContinue »